Site Notice | Privacy Policy | Downloads | Shop | Contact

Vacuum regulator Specht - plug-in units

COMPRESSED-AIR OPERATED

Medical vacuum regulator ejector is based on a compressed air ejector for regulation of suction;

with vacuum meter, fine control valve and quick action stop valve; Plug unit for direct connectionto a DIN outlet; With the fine control valve it is possible to regulate the vacuum infinitely.

All Woodpecker regulators meet DIN EN ISO 10079-3 specifications.

Questions?

Björn Rehr
+49 (0)40 / 73 93 57 - 0
E-Mail

Export customers:
E-Mail

  • TECHNICAL DATA
    Design:Vacuum generation using the ejector principle. Brass, chrome-plated.
    Medium:Compressed air (AIR)
    Inlet:Plug connector in accordance with DIN 13260-2 – gas-specific
    Primary pressure:450 kPa ± 100 kPa (in accordance with DIN EN ISO 7396-1)
    Suction inlet:9/16“-18 UNF with vacuum nozzle
    Dimensions (WxHxD) /Weight:110 x 160 x 110 mm / 700 g (Specht -90 / Specht -60)
    120 x 160 x 110 mm / 800 g (Specht -16)
    120 x 160 x 110 mm / 800 g (Specht -10)
  • PERFORMANCE PARAMETERS
     Suction adjustable betweenSuction flowrateCompressed air consumption
    Specht -900 and -90 kPa> 20 l/min
    at -90 kPa
    max. 30 l/min
    at -90 kPa
    Specht -600 and -60 kPa> 25 l/min
    at -60 kPa
    max. 30 l/min
    at -60 kPa
    Specht -160 and -16 kPa> 18 l/min
    at -16 kPa
    max. 20 l/min
    at -16 kPa
    Specht -100 and -10 kPa> 18 l/min
    at -10 kPa
    max. 15 l/min
    at -10 kPa

     

  • Order numbers
    Specht -90 vacuum regulator, AIR, 0 to -90 kPa, plug-in unit, DIN900.960
    Specht -60 vacuum regulator, AIR, to -60 kPa, plug-in unit, DIN901.590
    Specht -16 vacuum regulator, AIR, 0 to -16 kPa, plug-in unit, DIN900.958
    Specht -10 vacuum regulator, AIR, 0 to -10 kPa, plug-in unit, DIN901.914
our Retailers